Determination the content of ethanol in low alcohol beverages by gas chromatography
CSTR:
Author:
Affiliation:

(1. Xinjiang Academy of Analysis and Testing, Urumqi, Xinjiang 830011, China; 2. The Xinjiang Uygur Autonomous Region Occupation Disease Prevention and Control Center, Urumqi, Xinjiang 830011, China)

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    The ethanol content of beverages was measured by two methods, the static headspace gas chromatography and the traditional method of gas chromatography after distillation, using TG-WAXMS chromatographic column. The two methods had good linear relationship in the volume fraction of ethanol ranging from 0.01% to 5.00%. In traditional distillation method, the detection limit of volume fraction of ethanol was 0.003%, and the detection limit of static headspace was 0.001%. The results showed that due to the complicated pre-treatment processes, the recovery rate of the gas chromatography after traditional distillation was relatively low (83%~91%). While the method of direct determination of ethanol in the beverage by static headspace method was simple, and kept the ethanol stable during the detection. The recovery rate was between 97% and 100%.

    Reference
    Related
    Cited by
Get Citation

陈国通,毛琼玲,左书瑞,等.气相色谱法测定低酒精度饮料中乙醇的含量[J].食品与机械英文版,2018,34(2):58-61.

Copy
Related Videos

Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:December 13,2017
  • Revised:
  • Adopted:
  • Online: March 17,2023
  • Published:
Article QR Code